Microscope Facility
All of Hopkins

News & Status
2007_0713 Zeiss/Eppendorf Microinjection Station: Injector Repaired

2007_0706 Zeiss Meta #1: Recurring Software Problem

2007_0627 WBSB Room B13: Scheduled power outage (6/30/07-7/1/07)

2007_0625 Zeiss Meta #2 Upgrade: Scheduled for WED 06/27

2007_0618 Zeiss Meta #1 Available: Serviced by Zeiss

2007_0615 Zeiss Meta #1 Down: Problem with software

2007_0615 3i Spinning Disk Confocal: Fully Operational

2007_0612 3i Spinning Disk Confocal: Service Call Placed

2007_0612 Zeiss Meta #1: Available for limited use, service call placed

2007_0611 Zeiss Meta #2: Pinhole Alignment

2007_0607 Zeiss Meta #1: Serviced by Zeiss

2007_0606 Zeiss 510-Meta Confocals: New Policies Implemented

2007_0509 Zeiss Meta #1: Service Restored

2007_0509 Zeiss Meta #1 Down: Service Call Placed

2007_0508 Zeiss Meta #2: Pinhole Alignment

2007_0504 Zeiss Meta #2: Cy5 Configuration Corrected

2007_0503 Zeiss Meta #2: Cy5 Scans

2007_0502 3i SDC: Blue Excitation Laser Replaced

Staff Position Open &
Research Opportunities


Olympus IX81: Fluorescence Cube Selections

To accommodate User requests, we've changed the fluorescence cubes on the inverted microscope (Olympus Station 1). All prior fluorophores can still be visualized, but you may have to use different controls in IPLab to get to them.

1. If you do not need CFP or YFP, then you can select the fluorescence cube in the traditional way: (a) Click "OlympusIX-Log" and log "In" to the Olympus controller, (b) Click the "OlympusIX-Cube" button for a list of cube positions, and (c) Click the colored button next to the fluorophore-cube you want. If speed matters, (d) Click "OlympusIX-Log" and log "Out" of the Olympus controller.

Current "OlympusIX-Cube" choices: 1) ICG (infrared dye), 2) GFP, 3) TexasRed, 4) Cy5, 5) YFP/CFP (only works with 'Linking Device 1'; see below), 6) DIC

2. If you need CFP or YFP (or FRET between them), then you'll need to use a *different* interface: (a) Click "OlympusIX-Log" and log "In" to the Olympus controller, (b) Click the "Linking-Device 1" button for a list of cube positions "linked" to filter wheel settings, and (c) Click the fluorophore combination that you want. If speed matters, (d) Click "OlympusIX-Log" and log "Out" of the Olympus controller.

If you plan to switch between 'Linking-device 1' and 'OlympusIX-Cube' to control the dichroic filter turret, remember to leave 'Linking-device 1' with the filter 'wheels open', otherwise an extra emission/excitation filter is left in your optical path (probably would obscure your desired fluorescence). Because this is a complicated thing to remember, we've added extra settings under 'Linking-devices 1' automating the filter wheel to allow standard visualization of standard cubes other than YFP/CFP.

Current "Linking-Device 1" choices: 1) CFP, 2) YFP, 3) FRET, 4) wheels open, 5) Cy5, 6) Texas Red, 7) GFP, 8) ICG, 9) wheels open, 10) wheels open

Note: Linking devices 5-8 are offered as a convenience that duplicate "OlympusIX-Cube" settings (filter wheels set at 'open' positions).

 